U-Connectors for Blood-Contact Applications Made from Platinum-Cured Silicone Silicone U-Connectors

RAUMEDIC U-connectors are specially designed for use in blood-contact medical applications. Their U-shaped geometry, paired with high-purity silicone material, ensures gentle fluid transfer with minimal interaction and maximal safety for patients.

Request a Quote

Gentle and Biocompatible Silicone Connectors

The unique U-shape of the RAUMEDIC connectors ensures smooth, low-turbulence transitions in blood-contact applications. Manufactured from platinum-cured silicone and featuring a burr-free interior, they are exceptionally gentle on blood and tissue. Thanks to the inert properties of the silicone material, there is minimal interaction with fluids, and the absence of low-molecular compounds helps reduce the risk of extractables and leachables. These features make RAUMEDIC’s silicone U-connectors a trusted solution where safety, biocompatibility, and mechanical integrity are essential.

The connectors are produced under cleanroom conditions and are stored in standard warehouse stock, ensuring a high degree of supply security for our customers.

Product characteristics

  • Made from RAUMEDIC-SIK5683 platinum-cured silicone
  • Manufactured under ISO Class 7 cleanroom conditions
  • Biocompatible: meets ISO 10993-4/-5 and USP <85> endotoxin requirements
  • No surface printing; raised RAUMEDIC lettering according to internal drawing

Applications

  • Blood-contact connectors in extracorporeal circuits
  • Use with ECC silicone tubing systems
  • Suitable for heart-lung machines and critical care setups
Specifications
MaterialPlatinum-cured silicone rubber (RAUMEDIC-SIK5683)
Neutral color
Translucent to brownish-transparent depending on batch

Physical Properties

Specific Gravity: 1.13 g/cm³ (DIN EN ISO 1183-1A)
Hardness: 60 ± 5 Shore A (DIN EN ISO 868)
Tensile Strength: ≥ 7 MPa (DIN EN ISO 527)
Elongation at Break: 300–600% (DIN EN ISO 527)
Tear Resistance: ≥ 20 N/mm (ASTM D 624 B)
Burst Test: 2.0 kg/cm² (1.96 bar), no leakage after 10 min
Manufacturing EnvironmentManufactured in ISO Class 7 cleanroom (DIN EN ISO 14644)
Electrostatic charge may attract particles – absolute freedom from adherent particles cannot be guaranteed

Biocompatibility

Hemolysis tested (ISO 10993-4)
Cytotoxicity tested (ISO 10993-5)
Endotoxin tested (USP <85>, EP 2.6.14, FDA guideline)
Compliant with Ph. Eur. 3.1.9, FDA 21 CFR §177.2600, BfR XV
Sterilization MethodsEthylene Oxide (ETO)
Gamma Irradiation (up to 32 kGy)
Steam sterilization (up to 134°C)
Not suitable for repeated sterilization

Did we spark your interest? Request a Quote