U-Connectors for Blood-Contact Applications Made from Platinum-Cured Silicone Silicone U-Connectors
RAUMEDIC U-connectors are specially designed for use in blood-contact medical applications. Their U-shaped geometry, paired with high-purity silicone material, ensures gentle fluid transfer with minimal interaction and maximal safety for patients.
Gentle and Biocompatible Silicone Connectors
The unique U-shape of the RAUMEDIC connectors ensures smooth, low-turbulence transitions in blood-contact applications. Manufactured from platinum-cured silicone and featuring a burr-free interior, they are exceptionally gentle on blood and tissue. Thanks to the inert properties of the silicone material, there is minimal interaction with fluids, and the absence of low-molecular compounds helps reduce the risk of extractables and leachables. These features make RAUMEDIC’s silicone U-connectors a trusted solution where safety, biocompatibility, and mechanical integrity are essential.
The connectors are produced under cleanroom conditions and are stored in standard warehouse stock, ensuring a high degree of supply security for our customers.
Product characteristics
- Made from RAUMEDIC-SIK5683 platinum-cured silicone
- Manufactured under ISO Class 7 cleanroom conditions
- Biocompatible: meets ISO 10993-4/-5 and USP <85> endotoxin requirements
- No surface printing; raised RAUMEDIC lettering according to internal drawing
Applications
- Blood-contact connectors in extracorporeal circuits
- Use with ECC silicone tubing systems
- Suitable for heart-lung machines and critical care setups
| Material | Platinum-cured silicone rubber (RAUMEDIC-SIK5683) Neutral color Translucent to brownish-transparent depending on batch |
Physical Properties | Specific Gravity: 1.13 g/cm³ (DIN EN ISO 1183-1A) Hardness: 60 ± 5 Shore A (DIN EN ISO 868) Tensile Strength: ≥ 7 MPa (DIN EN ISO 527) Elongation at Break: 300–600% (DIN EN ISO 527) Tear Resistance: ≥ 20 N/mm (ASTM D 624 B) Burst Test: 2.0 kg/cm² (1.96 bar), no leakage after 10 min |
| Manufacturing Environment | Manufactured in ISO Class 7 cleanroom (DIN EN ISO 14644) Electrostatic charge may attract particles – absolute freedom from adherent particles cannot be guaranteed |
Biocompatibility | Hemolysis tested (ISO 10993-4) Cytotoxicity tested (ISO 10993-5) Endotoxin tested (USP <85>, EP 2.6.14, FDA guideline) Compliant with Ph. Eur. 3.1.9, FDA 21 CFR §177.2600, BfR XV |
| Sterilization Methods | Ethylene Oxide (ETO) Gamma Irradiation (up to 32 kGy) Steam sterilization (up to 134°C) Not suitable for repeated sterilization |